CIC Hard Drive Diagnostics Utility (Paid Version)

Supports the following Serial Numbered Equipment

  1. CIC ProTM  - Bedrock - Serial Number's starting with "JA"
  2. CIC ProTM  - Bedrock - Serial Number's starting with "SCH"
  3. CIC ProTM  - MP100 - Serial Number's starting with "SDY"
  4. CARESCAPE Central Station V1 - MP100 - Serial Number's starting with "SGJ
  5. CARESCAPE Central Station V2 - Serial Number's starting with "SKN"
  6. CTS - MP100 - Serial Number's starting with "SEE"

 

Usage

  1. Bedrock (Versions 4x & 5x)
    1. Insert USB Drive into an open USB slot on running Bedrock CIC.
    2. From the CIC Setup screen, select the Service Password tab.
    3. At the C:\Program Files\Marquette\CIC\4.0.x> prompt, type  “F:” and Enter
    4. In the Command prompt window type “1” and Enter.
    5. This will open a new window and then type “2” and Enter. This will launch the program.
    6. Select the Drive D: (Hard Drive) from the buttons under the Menu at the top.
      1.  NOTE: Drive C: and E: are the Compact Flash.
    7. Verify the Health Status
    8. Record the power on Hours in the spreadsheet if applicable.

 

  1. MP100 (Versions 5x & 6x), CARESCAPE Central Station V2 and CTS.​​
    1. At the C:\Program Files\Marquette\CIC\5.0.x> prompt, type  “H:” and Enter
    2. In the Command prompt window type “1” and Enter.
    3. This will open a new window and then type “2” and Enter. This will launch the program.
    4. Select the Drive G: from the buttons under the Menu at the top.
    5. Verify the Health Status
    6. Record the power on Hours in the spreadsheet if applicable.
Screen Shots

 CIC PRO / Bedrock

 

CarescapeTM MP100 / MP100R (Ver 5.x & 6.x), Carescape V2 & CTS​​

 

HD Spreadsheet

Data Entry Section

 

 Hard Drive by Hours