Carescape™ MP100D / MP100R Cooling Upgrade Kit
Clinical Technology Solutions is proud to offer the Cooling Upgrade Kit for the GE Healthcare Carescape Server. The upgrade will not only reduce the operating temperature of the unit, but increase the life of the unit and the up time. This cooling kit will reduce the cost of ownership of these units also. This is achieved by adding a simple fan to the unit with out any hardware modification.
Instructions:
- Shut down the Central station.
- Unplug the power cord from the unit.
- Open the MP100 case by removing all the case screws on the side.
- Vacuum and clean out the unit prior to installation if necessary,
- Locate the CPU Heatsink.
- Install the Fan on top of the Heatsink by pushing it on make sure you are square up to the heat sink.
- Plug in the fan to the CPU fan power located on the board between the CPU and memory.
- Replace the top cover.
- Power Caprescape™ MP100D / MP100R unit back on and you are done.
MP100 Running without CTS CPU Cooling (1 Minute) Temperature 67 C°
MP100 Running with CTS Cooling Kit (after 1 minute) Temperature 56 C° (11 C° Cooler than the CIC without Cooling Kit Fan)
MP100 Running with CTS Cooling Kit (after 10 minutes) Temperature 44 C° (23 C° Cooler the CIC without Cooling Kit Fan)