Drive Replacement Steps
- Shut down the CIC.
- Ground yourself on the chassis grounding point on the back of the CIC.
- Unplug the power cord from the back of the CIC.
- Remove the cover.
- With a #1 Phillips screwdriver, remove the four screws on each side of the CIC.
- It’s a good time to vacuum out the CIC and clear out any dust.
- Disconnect the SATA and power cables from the drive.
- With the Phillips screwdriver, loosen the four screws from the top hard drive and lift the drive straight up.
- Take the screws from the old drive and install them in the new drive.
- Install the new drive and tighten the screws snugly.
- Replace the cover and screws.
- Plug the CIC back in and power it up.
- Wait for the “System Settings Change” message and select Yes to reboot.
- You’re back up and running.
Supports the Following Serial-Numbered Equipment
CIC’s:
CARESCAPE Central Station V2 – serial “SKN”
CARESCAPE Central Station V1 / MP100 – serial “SGJ”
MP100R – serial “SEE”
MP100D – serial “SDY”
Bedrock – serial “SCH”
Bedrock – serial “JA”
BCM Tower CIC (SCSI HD)
Telemetry Servers:
MP100R – serial “SEE”
ATS – serial “SAH”
BCM Tower CIC / Telemetry Server (SCSI HD)
Keep Your Downtime Short