GE CIC / Telemetry Server Pre-Imaged Hard Drives

GE CIC / Telemetry Server Pre-Imaged Hard Drives

With our imaged hard drives, they will save you time and money.

  1. Reduce downtime.
  2. Reduce Repair time to ~15 minutes or less.

We offer both HDD and SSD Drives

FULL Disclosure Drives

  1. Shut down the CIC.
  2. Ground yourself on the chassis grounding point on the back of the CIC.
  3. Unplug the power cord from the back of the CIC.
  4. Remove the cover.
    1. With the #1 Philips screwdriver, remove the four screws on each side of the CIC.
    2. It is a great time to vacuum out CIC and eliminate some Dust Bunnies.
    3. Disconnect the SATA and Power Cables from the drive.
  5. With the Philips screwdriver, loosen the four screws from the TOP hard drive and lift the drive straight up.
  6. Take the screws from the old drive and install in the new drive.
  7. Install the new drive back in and tighten the screws snugly.
  8. Replace the cover and screws.
  9. Plug the CIC back in and Power up the CIC.
  10. Wait for the message “System Settings Change”  Select Yes to Reboot
  11. You are back up and running
Supports the following Serial Numbered Equipment
CIC’s:
  1. CARESCAPE Central Station V2 – Serial Numbers starting with “SKN
  2. CARESCAPE Central Station V1 – MP100 – Serial Numbers starting with “SGJ
  3. MP100R – Serial Numbers starting with “SEE
  4. MP100D – Serial Numbers starting with “SDY
  5. Bedrock – Serial Numbers starting with “SCH
  6. Bedrock – Serial Numbers starting with “JA
  7. BCM – Tower CIC (SCSI HD)
Telemetry Servers:
  1. MP100R – Serial Numbers starting with “SEE
  2. ATS – Serial Numbers starting with “SAH”
  3. BCM – Tower CIC / Telemetry Server (SCSI HD)